In electron spectroscopy, multi-channel detection combined with intense radiation sources provides the optimum experimental configuration. Building on the 5 mm, 192-channel ion detector developed at Aberystwyth, longer arrays have been fabricated for the detection of electrons in a commercial hemispherical analyser. The performance and reliability of a 10 mm, 384-detector array is discussed and the first array-detected photoelectron spectroscopy data for single-crystal diamond are presented. In scanning mode, the detector shows a large improvement compared to single channeltron detection and this improvement allows data to be collected in snapshot mode (1 s per spectrum) to enable real-time measurements.
| Iaith wreiddiol | Saesneg |
|---|---|
| Tudalennau (o-i) | 219-223 |
| Nifer y tudalennau | 5 |
| Cyfnodolyn | Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms |
| Cyfrol | 238 |
| Rhif cyhoeddi | 1-4 |
| Dynodwyr Gwrthrych Digidol (DOIs) | |
| Statws | Cyhoeddwyd - Awst 2005 |
Ôl bys
Gweld gwybodaeth am bynciau ymchwil 'A fully integrated multi-channel detector for electron spectroscopy'. Gyda’i gilydd, maen nhw’n ffurfio ôl bys unigryw.Dyfynnu hyn