Dose estimation using quartz OSL in the non-linear region the growth curve

A. S. Murray, A. G. Wintle, J. Wallinga

Allbwn ymchwil: Cyfraniad at gyfnodolynErthygladolygiad gan gymheiriaid

35 Dyfyniadau (Scopus)

Crynodeb

The reliability of equivalent-dose determinations on quartz made with the single-aliquot regenerative-dose (SAR) protocol is examined using a suite of sedimentary samples for which the equivalent dose is expected to be >200 Gy. Experiments on one sample were used to evaluate the precision obtained using the standard measurement procedure with (a) a range of pre-heat temperatures, and (b) a single pre-heat temperature. The precision is compared with that achieved using SAR for a repeated dose point after an initial pre-heat, and for a known dose given before the initial pre-heat. It is concluded that, at least for this sample, the scatter in OSL measurements dominates the spread in the estimates of De
Iaith wreiddiolSaesneg
Tudalennau (o-i)371-374
Nifer y tudalennau4
CyfnodolynRadiation Protection Dosimetry
Cyfrol101
Rhif cyhoeddi1-4
StatwsCyhoeddwyd - 2002

Ôl bys

Gweld gwybodaeth am bynciau ymchwil 'Dose estimation using quartz OSL in the non-linear region the growth curve'. Gyda’i gilydd, maen nhw’n ffurfio ôl bys unigryw.

Dyfynnu hyn