The calibration of any artificial β-source attached to a luminescence reader is fundamental for the accuracy of luminescence dating results. Here, we present calibration results obtained for a β-source attached to a single grain Risø reader in Bordeaux using a series of quartz of different origins. The quartz was irradiated with three different γ-irradiators. An unexpected variability of the apparent dose rates was observed and our results suggest that the γ-irradiation is the main reason for this variability. Further work is needed to clarify the underlying reasons.
Iaith wreiddiol | Saesneg |
---|---|
Tudalennau (o-i) | 1-10 |
Nifer y tudalennau | 10 |
Cyfnodolyn | Ancient TL |
Cyfrol | 37 |
Rhif cyhoeddi | 1 |
Statws | Cyhoeddwyd - 01 Meh 2019 |
Cyhoeddwyd yn allanol | Ie |