Crynodeb
This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the tupe Standard, hence designed for routine determination of palaeodoses using quartz of K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of De values determined using a Lexsyg reader with single grain De-values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%
| Iaith wreiddiol | Saesneg |
|---|---|
| Tudalennau (o-i) | 327-333 |
| Nifer y tudalennau | 7 |
| Cyfnodolyn | Geochronometria |
| Cyfrol | 41 |
| Rhif cyhoeddi | 4 |
| Dynodwyr Gwrthrych Digidol (DOIs) | |
| Statws | Cyhoeddwyd - 01 Rhag 2014 |
| Cyhoeddwyd yn allanol | Ie |
Ôl bys
Gweld gwybodaeth am bynciau ymchwil 'Performance tests using the Lexsyg luminescence reader'. Gyda’i gilydd, maen nhw’n ffurfio ôl bys unigryw.Dyfynnu hyn
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver