Performance tests using the Lexsyg luminescence reader

Johanna Lomax, Sebastian Kreutzer, Markus Fuchs

Research output: Contribution to journalArticlepeer-review

19 Citations (Scopus)

Abstract

This study presents test measurements on a Lexsyg luminescence reader, acquired by the Giessen luminescence group. The reader is of the tupe Standard, hence designed for routine determination of palaeodoses using quartz of K-feldspars. The tests include measurements of the stimulation powers, preheat temperatures, OSL- and TL-curves, as well as dose recovery tests and calibration measurements using highly sensitised calibration quartz. A comparison of De values determined using a Lexsyg reader with single grain De-values, which have been previously obtained on a Risø reader, is also presented. The results imply that the Lexsyg reader is a highly reliable measurement device with high reproducibility, yielding instrumental uncertainties of around 0.3%
Original languageEnglish
Pages (from-to)327-333
Number of pages7
JournalGeochronometria
Volume41
Issue number4
DOIs
Publication statusPublished - 01 Dec 2014
Externally publishedYes

Fingerprint

Dive into the research topics of 'Performance tests using the Lexsyg luminescence reader'. Together they form a unique fingerprint.