Projects per year
Abstract
Introduction
Flow infusion electrospray high resolution mass spectrometry (FIE-HRMS) fingerprinting produces complex, high dimensional data sets which require specialist in-silico software tools to process the data prior to analysis.
Objectives
Present spectral binning as a pragmatic approach to post-acquisition procession of FIE-HRMS metabolome fingerprinting data.
Methods
A spectral binning approach was developed that included the elimination of single scan m/z events, the binning of spectra and the averaging of spectra across the infusion profile. The modal accurate m/z was then extracted for each bin. This approach was assessed using four different biological matrices and a mix of 31 known chemical standards analysed by FIE-HRMS using an Exactive Orbitrap. Bin purity and centrality metrics were developed to objectively assess the distribution and position of accurate m/z within an individual bin respectively.
Results
The optimal spectral binning width was found to be 0.01 amu. 80.8% of the extracted accurate m/z matched to predicted ionisation products of the chemical standards mix were found to have an error of below 3 ppm. The open-source R package binneR was developed as a user friendly implementation of the approach. This was able to process 100 data files using 4 Central Processing Units (CPU) workers in only 55 seconds with a maximum memory usage of 1.36 GB.
Conclusion
Spectral binning is a fast and robust method for the post-acquisition processing of FIE-HRMS data. The open-source R package binneR allows users to efficiently process data from FIE-HRMS experiments with the resources available on a standard desktop computer.
Flow infusion electrospray high resolution mass spectrometry (FIE-HRMS) fingerprinting produces complex, high dimensional data sets which require specialist in-silico software tools to process the data prior to analysis.
Objectives
Present spectral binning as a pragmatic approach to post-acquisition procession of FIE-HRMS metabolome fingerprinting data.
Methods
A spectral binning approach was developed that included the elimination of single scan m/z events, the binning of spectra and the averaging of spectra across the infusion profile. The modal accurate m/z was then extracted for each bin. This approach was assessed using four different biological matrices and a mix of 31 known chemical standards analysed by FIE-HRMS using an Exactive Orbitrap. Bin purity and centrality metrics were developed to objectively assess the distribution and position of accurate m/z within an individual bin respectively.
Results
The optimal spectral binning width was found to be 0.01 amu. 80.8% of the extracted accurate m/z matched to predicted ionisation products of the chemical standards mix were found to have an error of below 3 ppm. The open-source R package binneR was developed as a user friendly implementation of the approach. This was able to process 100 data files using 4 Central Processing Units (CPU) workers in only 55 seconds with a maximum memory usage of 1.36 GB.
Conclusion
Spectral binning is a fast and robust method for the post-acquisition processing of FIE-HRMS data. The open-source R package binneR allows users to efficiently process data from FIE-HRMS experiments with the resources available on a standard desktop computer.
Original language | English |
---|---|
Article number | 64 |
Number of pages | 9 |
Journal | Metabolomics |
Volume | 18 |
Issue number | 8 |
DOIs | |
Publication status | Published - 02 Aug 2022 |
Keywords
- Mass spectrometry
- Metabolomic fingerprinting
- Post-acquisition processing
- Software
Fingerprint
Dive into the research topics of 'Spectral binning as an approach to post-acquisition processing of high resolution FIE-MS metabolome fingerprinting data'. Together they form a unique fingerprint.Projects
- 1 Finished
-
UK Consortium for MetAbolic Phenotyping (MAP UK)
Draper, J. (PI)
01 Jun 2019 → 31 May 2024
Project: Externally funded research